Hayes Handpiece is a provider of exceptional service. That is, we go beyond what others consider reasonable or even possible. It means:
- fast response times
- unrelenting improvement
- continuously adding value for you
- less downtime
- more control
- more information more quickly than ever before
In essence we eliminate or reduce:
- risk
- down time
- dollar operating cost of your handpieces each time you use them
We live by the same standards you set for your self. That is:
- We are trained, tested and independently certified to do the job
- We use specifically designed, original, high quality, safe tools and instruments
- We use genuine, certified materials and components
1. Schedule a free post bag
- Discuss the problem you are having
- FREE Express Post satchels in regional areas

2. Warranty Information
- Using our customised warranty tracking software, we determine if the handpiece is still covered under our repair warranty.
3. Clean and polish the external shell
- Test water spray, chuck strength and the turbine
- Test for air leaks
- Check fibre optics
- Look for any cracks or dents Ream head if applicable

4. Disassemble the handpiece if applicable
- Remove all moving parts and components.
5. Clean the handpiece
- Clear all components using proprietary cleaning techniques
- Remove all debris including rust from the drive air tube, exhaust tube and chip air and chip water tubes
- Clean internal head
- Clean head and remove dents when necessary
6. Check the impeller and spindle for wear and nicks
7. Test for concentricity
- Perform concentricity tests using custom made proprietary equipment called a concentricity gauge
- Maintain International Standards Organization (ISO) standards of .03 mm or less of total indicator run-out (TIR)


8. Overhaul handpiece and reassemble
- If the turbine is rebuildable, we rebuild it with certified manufacturers parts and bearings
- If the turbine failed chuck strength and/or concentricity tests, we install a new turbine
- Inspect and repair water spray to ensure proper atomization
- Check water and chip air tube
- Replace O-rings
- Replace fibre optics if necessary
- Replace end gasket if applicable
- Rethread if applicable
9. Test torque, RPM and noise

10. Deliver the handpiece and review maintenance procedures with your staff
- Discuss maintenance for high-speeds, low-speeds, swivels, scalers, fibre optics, and attachments
- Review different types of lubricants and cleaners
- Check air pressure for each handpiece
- Review different types of autoclaves and related procedures